Certifications
|
US EPA Federal Equivalent Method (FEM) for NO2 (0-500 ppb, 20-30°C),
EQNA-0217-243.
QAL1 Certified by TüV Rheinland for European standards EN15267 and EN 14211, certificate 0000062070.
|
Principle of Measurement
|
Direct absorbance of NO2 at 405 nm
|
Measurement Modes
|
NO2 only; NO only; NO2, NO and NOx
|
Linear Dynamic Range
|
0-10,000 ppb (0-10 ppm) for NO2; 0-2,000 ppb (0-2 ppm) for NO
(FEM approved for 0-500 ppb NO2 for 20-30°C)
|
Resolution
|
0.1 ppb
|
Precision (1σ rms noise)
|
< 0.5 ppb or 0.5% of reading (with adaptive filter1)
|
Accuracy
|
Greater of 2 ppb or 2% of reading
|
Limit of Detection (2σ)
|
< 1 ppb (with adaptive filter1)
|
Flow Rate (nominal)
|
1.5 Liter/min
|
Flow Rate Requirement
|
Minimum: 1.4 Liter/min; Maximum: 1.6 Liter/min
|
Response Time, 100% of Step Change
|
10 s for 5-s averaging
20 s with adaptive filter1
|
Measurement Frequency
|
0.2 Hz (once every 5 s)
|
Averaging Times
|
5 s, 1 min, 5 min, 1 hr
|
SD Card Logger Capacity
|
Minimum 2 GB (> 2-year capacity for 5-s measurement mode)
|
Internal Data Logger Capacity
|
8,192 data lines (5-s avg = 0.47 days; 1-min avg = 5.7 days; 5-min avg = 1.0 mo; 1-hr avg = 0.94 yr)
|
Concentration Units
|
ppb, ppm, pphm
|
Pressure Units
|
mbar
|
Temperature Units
|
°C
|
T and P Corrected
|
Yes
|
Operating Temperature Range
|
10 to 50°C (FEM approved for 20-30°C for NO2)
|
Power Requirement;
Supplied by 5-amp 110/220 VAC Power Pack (provided) or Battery
|
11-14 V dc or 120/240 V ac, 1.4 A at 12 V, 17 watt
Max: 2.9 A at 12 V, 35 watt (warmup)
|
Size
|
Rackmount: 17" w x 14.5" d x 5.5" h (43 x 37 x 14 cm)
|
Weight
|
18.6 lb (8.4 kg)
|
Data Outputs
|
RS232; 0-2.5 V Analog Outputs for NO and for NO2
|
Data Transfer Baud Rates
|
2400, 4800, 19200
|
Output Ranges
|
User-Defined Scaling Factor in Menu
|
DewLine?
|
Yes
|
Long-Life Pump
|
Yes: 15,000 hr, 1.7 yr of continuous operation
|
Flow Meter
|
Yes
|
Options
|
Bluetooth for wireless data transmission; USB output (in place of RS232)
|