Measurement Principle
|
UV Absorption at 254 nm
|
Linear Dynamic Range
|
0 to 2,000 μg/m3
|
Resolution
|
0.1 μg/m3
|
Precision (1σ rms noise)
|
Greater of 0.1 μg/m3 or 2% of reading (in 10-s measurement mode)
|
Limit of Detection (2σ)
|
0.2 μg/m3 (in 10-s measurement mode)
|
Accuracy
|
Greater of 0.1 μg/m3 or 2% of reading
|
NIST-Traceable Calibration
|
Yes
|
Measurement Interval
|
Fast mode: 2 s
Measurement mode: 10 s
Data averaging options: 1 min, 5 min, 1 hr
|
Flow Rate (nominal)
|
~0.8 Liter/min
|
Flow Rate Requirement
|
>0.5 Liter/min
|
Baseline Drift
|
< 0.1 μg/m3/day
< 5 μg/m3/year
|
Sensitivity Drift
|
< 1%/day
< 3%/year
|
Measurement Time, Frequency
|
Fast mode: 2 s, 0.5 Hz
Measurement mode: 10 s, 0.1 Hz
|
Response Time, 100% of Step Change
|
For 2-s output: 4 s, 2 points
For 10-s output: 20 s, 2 points
|
Averaging Times
|
1 min, 5 min, 1 hr
|
Data Storage
|
8,192 lines
(2-s fast mode ~4.6 hr; 10-s meas. mode ~1 day; 1-min avg ~6 days; 5-min avg ~1 mo; 1-hr avg ~1 yr)
|
Mercury Units
|
ppb, μg m-3
|
Pressure Units
|
torr
|
Temperature Units
|
K
|
T and P Corrected
|
Yes
|
Operating Temperature Range
|
0 to 50 °C
|
Operating Altitude Range
|
~0-13.5 km (150-1,013 mbar)
[modifications for higher altitudes available as an option]
|
Power Requirement;
Supplied by battery or 110/220 VAC Power Pack
|
7-24 V dc, nominally 250 mA at 12 V, 3.0 watt
|
External Battery
|
7.4 Volt, 1.6 amp hour, Lithium Ion Battery, 5-8 hr
|
Size
|
4.0 x 3.0 x 1.5 inches (10.2 x 7.6 x 3.8 cm)
|
Weight
|
0.75 lb. (0.34 kg)
|
Data Transfer Baud Rate
|
19200
|
DewLine?
|
Yes
|
GPS
|
Yes
|
Sampling Probe
|
Available as an option
|